\LARGE{Misplaced Trust in Software in the Therac-25}
\title{Misplaced Trust in Software in the Therac-25}
\author{Dan Ports}
\vspace{.1in}\paragraph{}

\end{center}
%\maketitle
%% The Therac-25's catastrophic failure was principally caused by
%% errors in the control software. In spite of the fact that software
%% interlocks were critical to ensuring the safety of the system, the
%% designers failed to consider the possibility of software
%% failures. When the first patient injuries were reported, the
%% manufacturer continued to focus on hardware malfunctions

The Therac-25's catastrophic failure was principally caused by errors
in the control software. The first patient injuries reported to the
manufacturer revealed the inadequacy of the software interlock system
for ensuring safe operation. The critical flaw was that the designers
did not take into account the possibility of software failures, in
spite of the fact that the system's safeguards were implemented
exclusively in software.

The Hamilton clinic incident, the first injury reported to the
manufacturer, resulted from the turntable being in the incorrect
position when the beam was activated. This unsafe condition should not
have been possible, and the fact that it occurred should have served
as a warning sign that the system design was deeply flawed. The system
should have had carefully validated interlocks in place to ensure that
unsafe conditions could not exist. Instead, the designers chose a
software system that was not sufficiently reliable.

AECL's engineers relied on software for safety-critical elements of
the Therac-25, and assumed the software was infallible. They
consistently failed to acknowledge that design or implementation
errors in the software could lead to safety problems. In the initial
design, they generated a fault tree, identifying potential ways in
which the hardware could fail. However, software errors were not
included in this fault tree, apparently because the engineers simply
assumed the software was correct and would never fail. The number of
software errors later encountered reveal the fallacy of this
thinking. In fact, this form of incomplete safety analysis may have
been more damaging than performing no safety analysis at all: the
misleading probabilities specified for hardware failures conveyed a
false sense of reliability, and led at least one user of the system to
assume that it must have been safe.

%% After the injury at the Hamilton clinic, the first one reported to
%% AECL, the manufacturer identified a microswitch in the turntable
%% assembly as a point of failure. Since the Therac-25 relied so heavily
%% on software, AECL's engineers ought to have thoroughly investigated
%% the software that made this malfunction possible. However, testing of
%% the software seems to have been neglected here. This oversight proved
%% tragic: the later Yakima incident, in which the same overdose occurred
%% despite the redesigned microswitch network, suggests that the failure
%% was, in fact, due to software error.

%% This particular response reflects a trend: the AECL engineers
%% continually ignored the possibility of software errors in their safety
%% analysis. They generated a fault tree during the design phase,
%% attempting to identify the probabilities with which each hardware
%% component would fail. However, software errors were not included in
%% this fault tree, apparently because they assumed the software was
%% correct and would not fail due to wear and tear. 

%% In fact, this form of
%% safety analysis may have been more damaging than performing no
%% analysis at all: 

Once problems began to appear, AECL focused on fixing specific bugs as
they became aware of them. This approach could not make the system
safe. Engineering a complex software system like the Therac-25's is no
simple task, and making it bug-free is practically impossible. As each
bug is fixed, it is only a matter of time until the next bug is
exposed, as was seen with the multiple bugs encountered in the
Therac-25. To develop a safety-critical system, it is necessary to
identify specific invariants that must be maintained: for example, the
beam must not be able to turn on if the turntable is not verified to be in
the correct position. These invariants must be maintained by carefully
tested safety interlocks. It is possible to rely on software, but only
if it has been very carefully tested and proven to ensure safe
conditions: the reused, inadequately-tested code in the Therac-25
certainly does not meet this standard. A set of redundant interlocks
is best; the Therac-20 caused no injuries despite probable software
flaws because a separate hardware safety shut down the system. If
these types of interlocks had been present on the Therac-25, the
tragic injuries may have been averted.